Risk of incorrect CAD-score

Incorrect placement of the CADScor® Sensor, outside patient IC4-L, may result in an incorrect CAD-score.

Inform patients who have CAD-score ≤20 to seek medical attention if symptoms persist or worsen after initial evaluation”, due to the risk of a false negative CADScor result.

The CAD-scores from patients in the 30-39-year group are currently outside the intended patient population. A warning triangle indicates the higher uncertainty of their CAD-score.